CHAPTER 2
Assembling Your Analyzer
13
To Connect:
1) Gently screw Microphone Type 4950 onto Preamplifier ZC-0032 (see items 1 and 2 in
Fig.2.1).
2) Insert the male plug of the preamplifier into the top socket of the Hand-held Analyzer (see
item 17 in Fig.2.1) and push gently until it snaps into position.
To Disconnect:
1) To remove the preamplifier and microphone, grip the locking collar firmly and slide
back, to remove the preamplifier and microphone combination from 2250 Light.
Mounting the Windscreen
For short outdoor noise measurements (or indoor measurements exposed to air movement),
mount Windscreen UA-0237 onto the microphone and preamplifier combination.
Mounting 2250 Light onto the Tripod Extension Stem and Tripod
Mount 2250 Light onto the Tripod Extension Stem UA-1651 and Small Tripod UA-0801, as
follows:
1) Screw Tripod Extension Stem UA-1651 onto the threaded stud of the ball-joint on Small
Tripod UA-0801. Secure the ball-joint in a roughly vertical position (that is, in-line with
the tripod) until ready to follow the instructions in step 4).
2) Screw the Extension Stem UA-1651 into the 2250 Light, using the threaded socket
situated on the underside of 2250 Light, at the back, (see item 18 in Fig.2.1).
3) Set Small Tripod UA-0801 (including 2250 Light) in the required position, and adjust it to
the required height. Ensure that one of the three legs is pointing in roughly the same
direction as your 2250 Light needs to point. (We will refer to this leg as the front leg.)
4) Position the extension stem at an angle of 45
° to the horizontal and vertically in-line with
the front leg of the tripod – this is to ensure the whole setup is stable.
Note: The procedure is the same if Tripod UA-0587 is used instead of Small Tripod UA-0801.
Once you have carried out these instructions, you are ready to start measuring, see Chapter 3.
2.5.3 Alternative Measurement Method (Extended Microphone)
The measurement microphone can be placed a distance from 2250 Light by connecting an
extension cable and fitting the microphone to a microphone holder. 2250 Light should be set up
using the following assembly instructions. This is to ensure that the accessories have limited
acoustical influence on the instrument during measurement. All the components described
below are shown in Fig.2.1 and Fig.2.4.