B&K 2250 Light Instrukcja Użytkownika Strona 141

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 140
CHAPTER 14
Specifications
133
Software Specifications – Logging Software for 2250 Light BZ-7133
The specifications for BZ-7133 include the specifica-
tions for Sound Level Meter Software for 2250 Light
BZ-7130. BZ-7133 adds:
Logging
MEASUREMENTS
Measurement data logged at pre-set periods into files
on:
SD Card: All hardware versions
CF Card: Hardware versions 1 – 3
USB Memory Stick: Hardware version 4
Logging Period: From 1 s to 24 hr with 1 s resolution
Fast Logging: L
AF
and L
Aeq
can be logged every
100 ms, irrespective of logging period
Broadband Data Stored at each Logging Interval:
All, or up to 10 selectable broadband data including
weather data and L
Aeq,T,mov
Broadband Statistics Stored at each Logging
Interval: Full distribution, or none
Spectrum Data Stored at each Logging Interval: All,
or up to three selectable spectra (license for BZ-7131
or BZ-7132 required)
Logging Time: From 1 s to 31 days with 1 s resolution
Measurement Total: For the logging time, in parallel
with logging: All broadband data, statistics and spectra
(license for BZ-7131 or BZ-7132 required)
Automatic reboot and resume of operation in case of
power failure
MEASUREMENT DISPLAYS
Profile: Graphical display of selectable measurement
data versus time. Fast display of next or previous
marker; Profile overview of entire measurement
Y-axis: Range: 5, 10, 20, 40, 60, 80, 100, 120, 140 or
160 dB. Auto zoom or auto scale available
X-axis: Scroll facilities
Cursor: Readout of measurement data at selected
time
MARKERS
Five user-definable markers for on-line marking of
noise sources or events anywhere in the profile.
Markers are set using the stylus on the touch screen,
or the three marker pushbuttons
Calibration
CHARGE INJECTION CALIBRATION (CIC)
Injects an internally generated electrical signal in
parallel with the microphone diaphragm. A manual CIC
can be performed whenever there is no measurement
in progress
An automatic CIC can be performed at the start and
end of a logging measurement
Storage
Measurement data is stored on an external memory
device.
For availability, please refer to the Ordering Infor-
mation
Notifications
ALARM CONDITIONS
CIC failed (in addition to those specified for BZ-7130)
Software Specifications – Signal Recording Option BZ-7226
Signal Recording Option BZ-7226 is enabled with a
separate license. It works with all the software for Type
2250-L: Sound Level Meter, Frequency Analysis and
Logging Software
For data storage, Signal Recording requires:
SD Card: All hardware versions
CF Card: Hardware versions 1 – 3
USB Memory Stick: Hardware version 4
RECORDED SIGNAL
A-, B-, C- or Z-weighted signal from the measurement
transducer
AUTOMATIC GAIN CONTROL
The average level of the signal is kept within a 40 dB
range, or the gain can be fixed
Przeglądanie stron 140
1 2 ... 136 137 138 139 140 141 142 143 144 145 146 ... 211 212

Komentarze do niniejszej Instrukcji

Brak uwag