Hand-held Analyzer 2250 Light – User Manual140
To compensate for this, use the built in windscreen correction.
A.2 Frequency Settings
Table A.2 Windscreen correction parameter
Parameter Values Comment
Windscreen Correction None
UA-0237
You can manually select a windscreen correction
for Windscreen UA-0237. No correction is made
for unknown microphones
Trigger Input None
MATRON
Handswitch
Voltage Level
This parameter should be set to match the
equipment connected to the Trigger Input socket
on the connector panel of the analyzer
Set it to None, if not used.
Set Trigger Input to MATRON Handswitch if the
analyzer is being used in the MATRON system.
This setting should also be used if you want to use
the Handswitch ZH-0680 for manually triggering
signal recordings
Set Trigger input to Voltage Level if you want to
control signal recording by a voltage level
generated by external equipment. The Voltage
Level should generate at least 2 V for On and less
than 1 V for Off. The duration of the steady level
should be at least 1 s, so it can be recognised by
the analyzer
Table A.3 Frequency Settings parameters
Parameter Values Comment
Broadband (excl. Peak) AC
AZ
BC
BZ
All broadband parameters (except L
peak
) are
measured simultaneously with two different
frequency weightings – select the weightings here
Broadband Peak X
C
Z
One broadband peak parameter L
peak
is
measured, select the frequency weighting here.
Note: X = frequency weighting A or B. ‘A’ requires
that the Broadband (excl. Peak) parameter is set to
AC or AZ. ‘B’ requires that the Broadband (excl.
Peak) parameter is set to BC or BZ