Configure CW Parameters
There are several parameters that should be set up prior to operating in CW mode. Press
so that it lights up, then press and to access Setup for CW mode. The
setup menu will be shown:
The setup parameters are: Range, High (Voltage limit), and Low (Voltage limit). Use the
keys to select each parameter, and use the numeric keypad to change the value. Press
to confirm the change.
Range
Use the numeric keypad to change the range. This value will also act as a limit to how much
power the load can be configured to. The adjustable ranges vary depending on the model. See
specifications for details.
High
This parameter refers to the voltage high limit for the automatic test mode. During automatic
test mode, the device under test (DUT) must be operating below the configured value for the
test to PASS upon completion. If the DUT operates above the configured value, the test will FAIL
upon completion.
Note: This parameter is used for Automatic Test Function ONLY.
Low