Transmitter Cond Ind 7100e/2(X)
86
Calibration solutions
Potassium chloride solutions
(Conductivity in mS/cm)
Temperature Concentration
*
[°C] 0.01 mol/l 0.1 mol/l 1 mol/l
0
5
10
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
0.776
0.896
1.020
1.147
1.173
1.199
1.225
1.251
1.278
1.305
1.332
1.359
1.386
1.413
1.441
1.468
1.496
1.524
1.552
1.581
1.609
1.638
1.667
1.696
7.15
8.22
9.33
10.48
10.72
10.95
11.19
11.43
11.67
11.91
12.15
12.39
12.64
12.88
13.13
13.37
13.62
13.87
14.12
14.37
14.62
14.88
15.13
15.39
15.64
65.41
74.14
83.19
92.52
94.41
96.31
98.22
100.14
102.07
104.00
105.94
107.89
109.84
111.80
113.77
115.74
1) Data source: K. H. Hellwege (Editor), H. Landolt, R. Börnstein: Zahlenwerte