Transmitter Cond Ind 7100e/2(X)
Passcodes according to FDA 21 CFR Part 11
Access to the device functions can be protected with adjustable
passcodes.
To call up the passcode editor:
Press conf key and enter Administrator passcode
(Factory setting: 1989).
Display Action
Remark
1. Press conf key.
2. Enter Administrator passcode
(1989):
Welcome text is displayed
This text is displayed
for approx. 3 sec
“Cal Info”
Edit: Arrow keys
Proceed with: enter
Cancel: conf
Default setting:
0000
“Zero adjustment”
Edit: Arrow keys
Proceed with: enter
Cancel: conf
Default setting:
1001
“Product calibration”
Edit: Arrow keys
Proceed with: enter
Cancel: conf
Default setting:
1105
“Temp probe adjustment”
Edit: Arrow keys
Proceed with: enter
Cancel: conf
Default setting:
1015
“Cell factor adjustment”
Edit: Arrow keys
Proceed with: enter
Cancel: conf
Default setting:
1100
“Adjustment with
entry of cal solution”
Edit: Arrow keys. Proceed with:
enter. Cancel: conf
Default setting:
0110
70098_E_CI7100_2XH.QXD 27.07.2005 9:27 Uhr Seite 56