Code
LIN
(LIN / LOG)
Display
Action
Choices
o1.
Select output characteristic
Select with arrow key
Proceed with enter
(Step omitted for % (Conc) or
SAL)
000.0 mS
(xxx.x mS)
100.0 mS
(xxx.x mS)
Conductivity
[mS/cm]
204
200
0
Assignment of measured values:
Current start and current end
[mS/cm]
204
200
100
[mA]
Example 1: Range 0...200 mS/cm
Example 2: Range 100...200 mS/cm
Advantage: Higher resolution in range
of interest
[mA]
Output current
Conductivity
With LIN selected:
Enter current start
(lower end of scale).
Select with key, edit number
with key, proceed with enter.
Enter current end
(upper end of scale).
Proceed with enter
Output current