Hand-held Analyzer Types 2250 and 2270 – User Manual
Fig.17.1 G-weighted frequency response
Low Frequency Option BZ-7234, together with BZ-7222, BZ-7223, BZ-7224 or BZ-7225 and
a low-frequency microphone, Type 4193 or Type 4964 (possibly with Low Frequency Adaptor
UC-0211, refer to Table C.5, Note for Type 4193 and Type 4964) measures the following timed
broadband parameters in addition to the standard measured parameters:
LGeq – G-weighted equivalent continuous noise level
LG10max – maximum time-weighted sound level measured with G-weighting and 10 s
exponential time constant
LG10min – minimum time-weighted sound level measured with G-weighting and 10 s
exponential time constant
LGpeak – maximum peak sound level with G-weighting
TGpeak – time when the G-weighted peak sound level occurred
The following instantaneous values are also available:
LG10 – instantaneous time-weighted sound level with G-weighting and 10 s exponential
time constant
LGpeak,1s – maximum peak sound level with G-weighting during the latest second
17.2 Setting up the Analyzer for G-weighting
1) Mount your low-frequency microphone and low-frequency adaptor (if used) on the
preamplifier and connect it to the top socket (see section 2.5.2). If it is your first time to
use the microphone, it needs to be created in the transducer database (see section 5.8) and
calibrated (see section 5.3).