CHAPTER 14
Reverberation Time Software (Optional Module)
161
14.4 Measurements with Graphical Position Management
The analyzer can keep track of the source and receiver positions in a graphical way together
with the measurements. To do this, make the following settings on the Setup screen:
1) Set Map Based Measurement = Yes in the Measurement Control setup – the reverberation
decays will then be placed graphically on the ‘map’ shown on the Overview tab, and
numbered as source-receiver relations (for example, S1R2 means the decay measurement
with noise from Source 1 and measured at Receiver position 2).
2) Set Meas. All Pos. for Each Source = Yes for measuring all combinations of source and
receiver positions. Set to No, if you want to measure at a specific number of receiver
positions per source.
3) Set No. of Positions per Source to the number of receiver positions you want to measure
per source.
4) Set Increment to Sources First, Receivers First or Manual. Typically Sources First is best
for Impulsive excitation (you can then walk between the source positions and generate the
impulses while the analyzer is placed on a tripod at one receiver position); Receivers First
is best for Interrupted Noise excitation (you can move the analyzer between the receiver
positions, while the sound source remains in the same place).
The Overview tab will then look like Fig.14.11:
Fig.14.11
Overview tab showing
one source and one
receiver position
Note the Add Source and Add Receiver icons on the headline for the map. Tap to
add new sources, and tap to add new receivers on the map.