Hand-held Analyzer Types 2250 and 2270 – User Manual
6) Steps 2 to 5 are automatically repeated a chosen number of times and the measured decays
are averaged together to reduce the uncertainty of the measurement.
7) The reverberation time spectra EDT, T20 and T30 are calculated and displayed on the screen.
The measurement can be in octaves or 1/3-octaves in parallel over a selectable frequency
range, allowing you to focus sound power on the relevant range. In each frequency band, the
decay is sampled 200 times each second, for reverberation times as long as 20 seconds.
You can enhance the signal-to-noise ratio by limiting the frequency range of the noise to cover
just a single frequency band, while keeping the output power constant. First, you make a
parallel measurement, then, if the signal-to-noise ratio is too low in a few bands (typically at
low frequencies) you select these bands for serial measurements. You can redo the
measurements including only these bands and merge the results with the parallel measurement.
Impulsive Excitation Method
With Impulsive Excitation, all you need to carry is Type 2250-F/2270-F, a tripod and a balloon
(or other impulsive source, such as a starting pistol). After you start the analyzer and pop the
balloon, the analyzer will start measuring, analyse the decay and display the RT spectrum and decay.
The single input range means that trial measurements are not necessary when using the impulsive
excitation method.
Fig.14.4
Typical RT measurement cycle for the impulsive excitation method
Fig.14.4 shows a typical RT measurement cycle performed by the analyzer, where the different
steps are described as follows:
1) After ‘Start’, the analyzer waits for the level to exceed the ‘Trigger Level’ (indicated on
the analyzer in three places by the “traffic light” giving a short green flash every second).
2) The impulse excitation is made, for example, a pistol is fired or a balloon is burst.
CAUTION: The use of hearing protection is highly recommended.
3) The impulse measurement is started 1 s before the level exceeds the ‘Trigger Level’.
Time
Trigger Level
Decay Time
Impulse measurement
1s
Await Level
to exceed
Trigger Level
Start
Impulse
Excitation
dB
Sound Level
060209