B&K 2250 Light Instrukcja Użytkownika Strona 10

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 16
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 9
10
BATTERY PACK
Rechargeable Li-Ion battery
Part No.: QB-0061
Voltage: 3.7 V
Capacity: 5200 mAh nominal
Typical Operating Time: >10 h (screen backlight dimmed); >7 h (full
screen backlight)
Battery Cycle Life: >500 complete charge/discharge cycles
Battery Indicator: Remaining battery capacity and expected working
time may be read out in % and time
Battery Fuel Gauge: The battery is equipped with a built-in fuel gauge,
which continuously measures and stores the actual battery capacity in
the battery unit
Charge Time: In instrument, typically 10 hours from empty at ambient
temperatures below 30°C. To protect the battery, charging will be
terminated completely at ambient temperatures above 40°C. At 30 to
40°C charging time will be prolonged. With External Charger ZG-0444
(optional accessory), typically five hours
Note: It is not recommended to charge the battery at temperatures below
0°C (32°F) or over 50°C (122°F). Doing this will reduce battery lifetime
CLOCK
Back-up battery powered clock. Drift <0.45 s per 24-hour period
Storage
INTERNAL FLASH-RAM (NON-VOLATILE)
20 Mbyte for user setups and measurement data
EXTERNAL MEMORY CARD
Secure Digital (SD and SDHC) Card: For store/recall of measurement
data
Compact Flash (CF) Card: Hardware versions 1 3 only. For
store/recall of measurement data
USB MEMORY STICK (Hardware Version 4 only)
For store/recall of measurement data
Environmental
WARM-UP TIME
From Power Off: <2 minutes
From Standby: <10 seconds
TEMPERATURE
IEC 60068–2–1 & IEC 6006822: Environmental Testing. Cold and
Dry Heat
Operating Temperature: –10 to +50°C (14 to 122°F), <0.1 dB
Storage Temperature: –25 to +70°C (–13 to +158°F)
HUMIDITY
IEC 60068–2–78: Damp Heat: 90% RH (non-condensing at 40°C
(104°F))
Effect of Humidity: < 0.1 dB for 0% < RH < 90% (at 40°C (104°F) and
1kHz)
MECHANICAL
Environmental Protection: IP 44
Non-operating:
IEC 60068–2–6: Vibration: 0.3 mm, 20 m/s
2
,
10 – 500 Hz
IEC 60068–2–27: Shock: 1000 m/s
2
IEC 60068–2–29: Bump: 4000 bumps at 400 m/s
2
WEIGHT AND DIMENSIONS
650 g (23 oz.) including rechargeable battery
300 93 50 mm (11.8 3.7 1.9) including preamplifier and
microphone
User Interface
LANGUAGE
User Interface in Catalan, Chinese (People’s Republic of China), Chinese
(Taiwan), Croatian, Czech, Danish, English, Flemish, French, German,
Hungarian, Japanese, Italian, Korean, Polish, Portuguese, Romanian,
Russian, Serbian, Slovenian, Spanish, Swedish and Turkish
HELP
Concise context-sensitive help in English, French, German, Italian,
Japanese, Korean, Polish, Portuguese, Romanian, Serbian, Slovenian
and Spanish
UPDATE OF SOFTWARE
Update to any version using BZ-5503 through USB or update via Internet:
any version from 4.0 and up
WEB PAGE
Connect to instrument via Internet browser supporting JavaScript. The
connection is password protected
Two Levels of Protection:
Guest level: for viewing only
Administrator level: for viewing and full control of the instrument
Software Specifications – Sound Level Meter Software for 2250 Light BZ-7130
Conforms with the following National and International Standards:
IEC 616721 (200205) Class 1
IEC 60651 (1979) plus Amendment 1 (1993 02) and Amendment 2
(200010), Type 1
IEC 60804 (200010 ), Type 1
DIN 45657 (199707)
ANSI S1.41983 plus ANSI S1.4 A1985 Amendment, Type 1
ANSI S1.431997, Type 1
Note: The International IEC Standards are adopted as European
standards by CENELEC. When this happens, the letters IEC are replaced
with EN and the number is retained. 2250 Light also conforms to these
EN Standards
Input
CORRECTION FILTERS
For Microphone Type 4950 and 4952:
Correct the frequency response to compensate for sound field and
accessories
Accessories (Type 4950 only): None or Windscreen UA-0237
Sound Field: Free-field or Diffuse-field (for Type 4952 only: 0° (Top)
reference direction and 90° (Side) reference direction)
Analysis
DETECTORS
Parallel Detectors: On every measurement:
A- or B-weighted: (Switchable) broadband detector channel with three
exponential time weightings (Fast, Slow, Impulse), one linearly averaging
detector and one peak detector
C- or Z-weighted: (Switchable) as for A- or B-weighted
Overload Detector: Monitors the overload outputs of all the frequency
weighted channels
MEASUREMENTS
X = frequency weightings A or B
Y = frequency weightings C or Z
V = frequency weightings A, B, C or Z
U = time weightings F or S
Q = exchange rate 4, 5 or 6 dB
N = number between 0.1 and 99.9
For Storage: Full statistics